Yamamoto, Masahiko; Horigome, Kazushi; Kuno, Takehiko
Applied Radiation and Isotopes, 190, p.110460_1 - 110460_7, 2022/12
Times Cited Count:1 Percentile:29.26(Chemistry, Inorganic & Nuclear)Gravimetric measurement of U content in UO with ignition in the air has been investigated. The ignition temperature, ignition time and aliquot sample mass are optimized as 900C, 60 minutes, and 1 g, respectively. The method is validated by IDMS with uncertainty estimation. The obtained result by gravimetry is 0.782360.00051 g/g (k=2) and agreed with IDMS value within its uncertainty. It has been found that U in UO can be measured accurately and precisely by gravimetry.